DEPUY ORTHOPAEDICS INC US SUMMIT BASIC PRESS FIT SZ 3; SUMMIT HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1570-05-090
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Limb Fracture (4518)
Date of Event 12/30/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Patient reportedly fell causing a periprosthetic fracture.Stem and modular cathcart hip ball were removed and replaced with revision corail stem and new hip ball.Doi: (b)(6) 2022.Dor: (b)(6) 2022.Affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT BASIC PRESS FIT SZ 3
Common Device Name
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JJM (SUZHOU) LTD. 3006356043
no.299 changyang st
suzhou industrial park
suzhou jiangsu 21512 -6
CH   21512-6
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16145953
Report Number1818910-2023-01055
Device Sequence Number20272169
Product Code LPH
UDI-Device Identifier10603295059691
UDI-Public10603295059691
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K170339
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1570-05-090
Device Catalogue Number157005090
Device Lot NumberD20061096
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2022
Supplement Date Received by Manufacturer01/27/2023
Initial Report FDA Received Date01/12/2023
Supplement Report FDA Received Date01/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MODULAR CATHCART BALL 48MM OD; TAPERED SPACER ARTICUL/EZE -3
Outcome Attributed to Adverse Event Required Intervention;
Patient Age89 YR