LIVANOVA DEUTSCHLAND HEATER COOLER 3T (230 VOLT); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-80 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/15/2022 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that, during service, a heater-cooler system 3t displayed error code (e6) associated to pump using too much power in the display of the patient tank.There was not patient involvement.
|
|
Manufacturer Narrative
|
There was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in united kingdom.The livanova field service representative in charge found the circulation motor was jammed and would not turn, resistors were burnt out on distribution board.The circulation motor and distribution board were replaced and this did not solve the issue: error e06 reappeared.The cpu was replaced and full calibration performed.This solved the issue: the error cleared and heater-cooler began to function properly.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
A complaints database review has been performed and no further similar issue has been submitted for this unit since its manufacturing date in 2021.Based on the above facts, the root cause of the reported error code can be traced back to defective boards and a jammed circulator motor.It cannot be ruled out water infiltration could have led to the reported damage on the electronics of the board and the motor.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
|
Event Description
|
See initial report.
|
|