SYNTHES GMBH 10MM/125 DEG TI CANN TFNA 340MM/LEFT - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Model Number |
04.037.025S |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from japan reports an event as follows: it was reported that on (b)(6) 2022, the patient underwent orif for subtrochanteric fracture with the products in question.The patient was a young male with good bone quality.It is possible that the nail was slightly shaved and the residual iron debris in the body, because the guide pin of the blade has been inserted in a squished position and when reaming the bone head the squished guide pin may have misled causing the bone head reamer to interfere with the nail.The surgery was completed without any surgical delay.Postoperatively, the guide pin and osteoclavicular reamer were checked, and since they were not shaved, it is likely that the nail was shaved.No further information is available.This report is for a 10mm/125 deg ti cann tfna 340mm/left ¿ sterile.This is report 1 of 2 for (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|