| Model Number |
2054 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.Device not returned.
|
| |
|
Event or Problem Description
|
|
The customer reported that the sensor suddenly stopped working.No consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing. the sensor was determined to be functioning as designed.
|
| |
|
Event or Problem Description
|
|
The customer reported that the sensor suddenly stopped working.No consequences or impact to patient were reported.
|
| |
|
Search Alerts/Recalls
|