• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. DISPOSABLE ELECTROSURGICAL SNARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. DISPOSABLE ELECTROSURGICAL SNARE Back to Search Results
Model Number SD-230U-20
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was returned to olympus for evaluation without the original packaging.During inspection and testing, the allegation was confirmed.A visual inspection found foreign substance adhered to the loop tip and debris on the loop wires.There was a charring substance and biomaterial stains inside the distal tube sheath as signs of use.However, no physical damage was found on the tube sheath, and the handle slider was functioning properly.The device passed the continuity check between the loop and plug contact pin.Based on the evaluation findings, the reported complaint was confirmed; however, the device had signs of being used instead of an out of box failure.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event or Problem Description
The customer reported to olympus, tissue found on snare out the box of a disposable electrosurgical snare.The event occurred during preparation for use.The doctor noticed foreign substance on the snare before passing it down the scope and as a result did not use the device on a patient.A new device was opened with no issues.There was no reported patient harm associated with this event.
 
Event or Problem Description
Additional information was provided regarding this event.The intended procedure was an unspecified therapeutic procedure.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the foreign material could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE ELECTROSURGICAL SNARE
Common Device Name
DISPOSABLE ELECTROSURGICAL SNARE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16153612
Report Number9614641-2023-00049
Device Sequence Number7507247
Product Code FDI
UDI-Device Identifier04953170328190
UDI-Public04953170328190
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K955650
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSD-230U-20
Device Lot NumberV2823
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2022
Supplement Date Received by Manufacturer01/17/2023
02/20/2023
Initial Report FDA Received Date01/12/2023
Supplement Report FDA Received Date02/16/2023
03/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-