| Model Number |
VTICM5_13.2 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Type of investigation - lens work order search: no similar complaint type event reported for units within the same lot.Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 13.2mm, vticm5_13.2, -12.5/4.0/73 (sphere/cylinder/axis), implantable collamer lens was implanted into the patients left eye (os) on (b)(6) 2022 and the patient experienced excessive vault, lens remains implanted.Cause of event was unknown.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5- "on (b)(6) 2023 the lens was exchanged for the same model/shorter length lens and the problem was resolved.Status of the eye was "all fine".Additional information provided was "patient is happy".Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|