ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number SSLP-235
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 11/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii/iv.
 
Event or Problem Description
Healthcare professional reported left side capsular contracture baker grade iii/iv.Device remains implanted.
 
Event or Problem Description
Healthcare professional reported left side capsular contracture baker grade iii/iv.Device has been explanted and replaced.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ capsular contracture: unable to observe since it is a medical event and is not related to the device.Additional observations: ¿ deformation is observed.¿ voids is observed.No further actions are required as the device was implanted.
 
Event or Problem Description
Healthcare professional reported left side capsular contracture baker grade iii/iv.Device has been explanted and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key16157682
Report Number9617229-2023-00464
Device Sequence Number17991101
Product Code FTR
UDI-Device Identifier10888628032699
UDI-Public10888628032699
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/06/2022
Device Model NumberSSLP-235
Device Catalogue NumberSSLP-235
Device Lot Number3124737
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/29/2022
Supplement Date Received by Manufacturer04/26/2023
07/12/2023
Initial Report FDA Received Date01/13/2023
Supplement Report FDA Received Date05/15/2023
08/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/06/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
Patient SexFemale
Patient Weight55 KG
-
-