| Model Number |
101/860/085CZ |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the pre-use check, a pinhole was found in the cuff.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device serial number/lot number is unknown, no product information has been provided to date.A product sample was received and is awaiting evaluation, investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: one unit was returned for investigation.Under visual inspection the sample appeared to be in good condition.Customer is claiming that the failure was observed prior pre-use check but there were found some biological residues on tracheostomy tube which indicates that product was in use.It was found that sample cannot pass this 12 hour test - cuff was deflating during inflation.Source of leak was found to be tear in cuff.It is the most probable that reported failure occurred during tracheostomy procedure or during use due to contact with a sharp edge.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of product.
|
| |
|