| Model Number |
WA02946A |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation and the customer¿s allegation was confirmed.It was noted that there were black spots on the image.The outer tube was bent and was missing the inner outer adapters.There was debris under the distal end cover glass and dents were noted on the frame.Debris was also seen under the optical system window and scratches were noted on the funnel.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
It was reported to olympus, that the autoclavable ureteroscope had a cracked lens.There were no reports of patient harm associated with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be determined.It is likely the event occurred due to the use of excessive force by the customer.Olympus will continue to monitor field performance for this device.
|
| |
|