| Model Number |
OPT318 |
| Medical Device Problem Codes |
Break (1069); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A distributor in china on behalf of a healthcare facility reported that tubing of an opt318 optiflow junior nasal cannula was cracked during patient use.There was no patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Fisher & paykel healthcare (f&p) are currently in the process of retrieving further information regarding this complaint.We will provide a follow up report upon the completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A distributor in china on behalf of a healthcare facility reported that tubing of an opt318 optiflow junior nasal cannula was cracked during patient use.There was no patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
Ps412521.The optiflow junior nasal cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the complaint opt318 optiflow junior nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the picture and information provided by the customer and our knowledge of the product.Results: visual inspection of the provided photograph revealed that there were four holes in one tubing.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.All optiflow junior cannula are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are taken hourly from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint.If there are any failures the entire batch is put aside for further investigation.The subject opt318 optiflow junior nasal cannula would have met the required specification at the time of production.The user instructions which accompany the opt318 optiflow junior nasal cannula show in pictorial format the correct placement and fitting of the cannula and also warn: ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.Appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death.Do not stretch or crush tube.
|
| |
|