This device is classified as import for export; therefore, 510k is not applicable.Model: eg-1690k is available in the usa with a 510k number: k131902.We checked the returned unit and confirmed that the air/water channel clogged.Based on the result, we concluded that it was caused due to the insufficient reprocessing at the facility on the air/water channel.In addition, we confirmed that the control body fluid damage, the u/d pulley wire fluid damage, the r/l pulley wire fluid damage, the light guide fiber bundle (lcb) broken, the operation channel buckled, and the air/water nozzle clogged; however, they are not the main cause and/or irrelevant to the alleged complaint.Based on the technical report, "hr-rpt-0630(air/water & jet water channels)" and/or the risk analysis results, it was evaluated to submit mdr.
|