| Model Number |
02.18.IF3.08.RM |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 03 january 2023: lot 2108470: (b)(4) items manufactured and released on 27-sept-2021.Expiration date: 2026-09-07.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event in the period of review.
|
| |
|
Event or Problem Description
|
|
The patient came in for a post-op appointment and it was observed that patient's femoral component would impinge upon cement in flexion.At about 8 months post primary the surgeon removed the poly, cleared out cement posteriorly and put in a new poly successfully.
|
| |
|