|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d9, g3, g6, h1, h2, h3, h6, h10.Visual examination of the returned product identified slight wear and discolouration on the top articulating radius consistent with being implanted for more than 5 years, otherwise the bearing appears to be in good condition.A review of the device manufacturing records confirmed no abnormalities or deviations.Device used for treatment.Radiographs were provided and reviewed from an radiologist: significant findings include subluxation of the femoral component as well as displaced polyethylene lining on the later time point.The original time points demonstrate possible improper sizing of the hardware for the bones.With the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|