ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
|
Back to Search Results |
|
Model Number B12LT |
Device Problems
Crack (1135); Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Date of event: unknown, assumed first day of month that complaint was reported.Batch #: unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
It was reported that when the lobectomy surgery was finished, the trocar was noted to be broken when taking it out.All the pieces were retrieved.No patient consequences.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 3/14/2023.D4: batch # unk.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that only the sleeve from the b12lt device was received with the sleeve broken and with the duckbill out of position.In addition, the duckbill and universal seal were not returned.No conclusion could be reached regarding what may have caused the reported incident.One possible cause for the damage found may be due to excessive force being applied to the device.Please refer to the instructions for use for additional information regarding proper device usage.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.The manufacturing records could not be reviewed as the batch number is unknown.H6.Medical device problem code: a0404.
|
|
Search Alerts/Recalls
|
|
|