| Catalog Number |
1620002 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hernia (2240)
|
| Date of Event |
06/20/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This legal event is being reported as serious injury due to the reported recurrence with surgical intervention.The internal investigation is pending and will be reported in a follow up report along with the investigation conclusion.
|
| |
|
Event or Problem Description
|
|
It was reported through a legal event that a 38 year old female had hernia repair surgery on (b)(6) 2012.During the hernia repair surgery, the surgeon implanted a strattice mesh.The lot and batch number for this mesh is s1106061-172.After surgery, the patient returned to the hospital on or about april 19, 2013.She was diagnosed with recurrent ventral hernia.The remaining strattice mesh was removed and additional mesh was implanted.On or about january 26, 2015, the patient presented with a recurrent incarcerated ventral incisional hernia and additional strattice was implanted.She underwent a series of additional revision surgeries, with strattice being removed on or around (b)(6) 2016.On or about (b)(6) 2021 the patient presented with a recurrent ventral hernia that was repaired with strattice.The catalog number for this mesh is 1620002 and the lot number is sp200133-040.No other information was reported.This record is associated with the device implanted (b)(6) 2015; lot unknown.Although it was reported there was placement of an additional mesh during the april 19, 2013, the record does not indicate strattice, therefore a record will not be opened for this mesh.The patient was reported to be implanted with a 3rd strattice mesh on (b)(6) 2021; however there is no event reported after the date of implant; therefore no additional record will be opened.
|
| |
|
Event or Problem Description
|
|
This is follow up#1 to report the results from the internal investigation and the conclusion.As reported in the initial: it was reported through a legal event that a 38 year old female had hernia repair surgery on (b)(6) 2012.During the hernia repair surgery, the surgeon implanted a strattice mesh.The lot and batch number for this mesh is s1106061-172.After surgery, the patient returned to the hospital on or about (b)(6) 2013.She was diagnosed with recurrent ventral hernia.The remaining strattice mesh was removed and additional mesh was implanted.On or about (b)(6) 2015, the patient presented with a recurrent incarcerated ventral incisional hernia and additional strattice was implanted.She underwent a series of additional revision surgeries, with strattice being removed on or around (b)(6) 2016.On or about (b)(6) 2021, the patient presented with a recurrent ventral hernia that was repaired with strattice.The catalog number for this mesh is 1620002 and the lot number is sp200133-040.No other information was reported.This record is associated with the device implanted (b)(6) 2015; lot unknown.Although it was reported there was placement of an additional mesh during the (b)(6) 2013, the record does not indicate strattice, therefore a record will not be opened for this mesh.The patient was reported to be implanted with a 3rd strattice mesh on (b)(6) 2021; however there is no event reported after the date of implant; therefore no additional record will be opened.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal investigation into strattice lot sp200133 included a review of the reported information, review of the device history records, and a review of the complaint history records.The investigation resulted in no remarkable findings, including no other complaints reported against the lot and no deviations or related non-conformances revealed during processing.The lot was terminally sterilized within the process parameters and met all qc release criteria.As of 19 jan 2023, of the (b)(4) devices released to finished goods for lot sp200133, 93 have been distributed with 41 reported as implanted.Based on our internal investigation with no remarkable findings, and without relevant patient factors, a relationship between the strattice and this event could not be determined.Due to the legal process, if additional information is made available during legal proceedings, a supplemental report will be submitted.As reported in the initial: this legal event is being reported as serious injury due to the reported recurrence with surgical intervention.The internal investigation is pending and will be reported in a follow up report along with the investigation conclusion.
|
| |
|
Search Alerts/Recalls
|
|
|