• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS SYMBOTEX; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS SYMBOTEX; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number SYM20
Device Problems Mechanics Altered (2984); Adverse Event Without Identified Device or Use Problem (2993); Material Split, Cut or Torn (4008)
Patient Problems Adhesion(s) (1695); Emotional Changes (1831); Inflammation (1932); Nausea (1970); Pain (1994); Scar Tissue (2060); Seroma (2069); Vomiting (2144); Hernia (2240); Distress (2329); Obstruction/Occlusion (2422); Abdominal Distention (2601); Unspecified Tissue Injury (4559)
Event Type  Injury  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The patient's attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an incisional hernia.It was reported that after implant, the patient experienced defective mesh, hole in mesh, pain, emotional distress, disability, impairment, loss of enjoyment of life, mental pain, recurrence, obstruction, adhesions, serosal tear, protrusion of colon, inflammation.Post-operative patient treatment included revision surgery, ct scan, takedown of adhesions, hernia repair.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The patient's attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an incisional hernia.It was reported that after implant, the patient experienced suffering, defective mesh, hole in mesh, pain, emotional distress, disability, impairment, loss of enjoyment of life, mental pain, recurrence, obstruction, adhesions, serosal tear, protrusion of colon and inflammation.Post-operative patient treatment included hernia repair with additional mesh, serosal tear repair, exploratory laparotomy, revision surgery, ct scan, takedown of adhesions and ventral hernia repair.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional info: a1, b2 (added hospitalization), h6 (patient codes, imf code).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
The patient's attorney alleged a deficiency against the device.The product was used for therapeutic laparoscopic treatment of an incisional hernia.It was reported that after implant, the patient experienced scarring, scar tissue, seroma, nausea, vomiting, stranding around loop of bowel, suffering, defective mesh, hole in mesh, pain, emotional distress, disability, impairment, loss of enjoyment of life, mental pain, recurrent hernia with incarceration, obstruction, adhesions, serosal tear, protrusion of colon and inflammation.Post-operative patient treatment included hospitalization, scar revision, application of prevena dressing, hernia repair with additional mesh, serosal tear repair, exploratory laparotomy, revision surgery, ct scan, takedown of adhesions and ventral hernia repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMBOTEX
Type of Device
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16185301
MDR Text Key307494686
Report Number9615742-2023-00061
Device Sequence Number1
Product Code FTL
UDI-Device Identifier10884521593435
UDI-Public10884521593435
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142908
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/17/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSYM20
Device Catalogue NumberSYM20
Was Device Available for Evaluation? No
Date Manufacturer Received02/15/2023
Date Device Manufactured12/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN PROTACK.
Patient Outcome(s) Hospitalization; Required Intervention; Disability;
Patient SexMale
-
-