| Model Number |
WA50372B |
| Medical Device Problem Codes |
Break (1069); Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A telescope (laparoscope) was returned to olympus for repair due to a damaged eyepiece.Image failure due to lens damage was reported.It is unknown when the reported issue was found.There was no harm or user injury reported due to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus.The investigation is still in progress; therefore, the root cause of the reported defect cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.Olympus will continue to monitor the field performance of this device.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was returned to olympus for inspection, and the customer's complaint was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be determined.It is likely the reported events, lens and eyepiece damage, occurred due to the use of excessive force.Olympus will continue to monitor field performance for this device.
|
| |
|