| Model Number |
629383S |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Nonunion/Delayed-union Bone Fracture (2369); Implant Pain (4561)
|
| Date of Event |
12/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported: "patient complains of pain and bone fusion failure and screw backout were found.It was determined that there was no infection, but there was a suspicion of metal allergy.".
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "patient complains of pain and bone fusion failure and screw backout were found.It was determined that there was no infection, but there was a suspicion of metal allergy.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, since an x-ray was provided which matches the alleged failure mode.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.An x-ray was provided which was presented to our health care professional for evaluation, and a question about the most likely reason for the screw backout was asked, to which the medical expert replied: ¿with the limited information available it is almost impossible to give a root cause.The first cause could be improper locking of the locking screws, or improper use of the locking screws (outside of the cone/ angle in which the screws can be properly locked.Second could be related to the construction being unstable, which may have caused too much movement of the parts, which lead to screw movement and loosening.However as said, based on the available information an actual cause cannot be determined.¿ more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
|
| |
|