• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G6 PROFESSIONAL SYSTEM; CONTINUOUS GLUCOSE MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G6 PROFESSIONAL SYSTEM; CONTINUOUS GLUCOSE MONITOR Back to Search Results
Model Number 9445-20
Medical Device Problem Code No Device Output (1435)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that transmitter failed error occurred.Data was evaluated and the allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G6 PROFESSIONAL SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key16190699
Report Number3004753838-2023-013909
Device Sequence Number4338334
Product Code QII
UDI-Device Identifier00386270001092
UDI-Public00386270001092
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K191833
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date07/11/2023
Device Model Number9445-20
Device Catalogue NumberSTP-PR-001
Device Lot Number5314665
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/19/2022
Initial Report FDA Received Date01/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/24/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age35 YR
Patient SexFemale
-
-