|
|
| Model Number |
VNMF3428C173TU |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical devices: other relevant device(s) are: product id: v nmc3428c207tu, serial/lot #: (b)(4), ubd: 23-sep-2021, udi#: (b)(6); product id: vnmc3434c90tu, serial/lot #: (b)(4), ubd: 19-aug-2021, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Valiant navion stent grafts were implanted during the endovascular treatment of a type b thoracic dissection.It was reported by the patients wife reported that the patient has undergone an additional vascular surgery approximately 2 years post the index procedure.Cta findings from approximately 32 months post the index procedure showed false lumen perfusion.Compared to ct examination 26 months post the index procedure shows there remains endoleak's about the aortoiliac bifurcation.The native aorta about this level remains relatively unchanged in dimension, currently measuring 51 x 47 mm in the transverse and ap dimensions respectively.There are additional endoleak's about the distal descending thoracic aorta at the level of the diaphragmatic crus.Left common iliac-sma, and left common iliac-bilateral renal artery debranched bypass grafts remain widely patent.There is good visceral perfusion of the bowel, liver, and bilateral kidneys from these bypass grafts.Per the physician the cause of the false lumen and endoleaks are undetermined.No additional clinical sequelae were reported and the patient will be monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
B5; additional information received: it was confirmed the subtype of the endoleaks seen at the aortoiliac bifurcation and distal descending thoracic aorta were type i endoleaks.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|