MAUDE Adverse Event Report: PHILIPS ELECTRONICS NEDERLAND B.V. INGENIA 1.5T

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ELECTRONICS NEDERLAND B.V. INGENIA 1.5T Back to Search Results
Model Number INGENIA 1.5T
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Partial thickness (Second Degree) Burn (2694)
Date of Event 11/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Philips received a report that a patient sustained a 2nd degree burn with skin damage after a lumbar spin examination.
 
Additional Manufacturer Narrative
Conclusion.Based on the available information, there is no indication of a malfunction of the mr system or coil used that could have contributed to the incident.The observed reddening and skin damage can be explained by close or direct contact between the patient and the bore wall.No padding was used to prevent this.Contributing factors: the thermoregulation of obese and hypertensive patients is known to be impaired.The risk of rf energy-related injuries is higher in patients with impaired thermoregulation.Several scans with high sar values (>2 w/kg) were executed.High sar scans are a bigger challenge for the cool down mechanism than low sar scans.The total administered specific energy dose of 3.7 kj/kg is very high for a patient with impaired thermoregulation.The instructions for use advises to avoid sed >3.0 kj/kg, preferably keep sed <2.0 kj/kg for those patients.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGENIA 1.5T
Common Device Name
INGENIA 1.5T
Manufacturer (Section D)
PHILIPS ELECTRONICS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS ELECTRONICS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
222 jacobs street
cambridge, MA 02141
MDR Report Key16193690
Report Number3003768277-2023-00340
Device Sequence Number17214775
Product Code LNH
UDI-Device Identifier00884838009820
UDI-Public884838009820
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K110151
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberINGENIA 1.5T
Device Catalogue Number781396
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2023
Supplement Date Received by Manufacturer01/10/2023
Initial Report FDA Received Date01/18/2023
Supplement Report FDA Received Date03/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age64 YR
Patient SexMale
Patient Weight162 KG
-
-