• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. BASX BLADELESS TROCAR 12MM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. BASX BLADELESS TROCAR 12MM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number TB12LT
Medical Device Problem Codes Component or Accessory Incompatibility (2897); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # unk.An analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a gastric bypass, the trocar seal was leaking.The leakage is caused by the seal being damaged from the endo-gia stapler being removed through the trocar.Procedure prolonged 5 minutes.No patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 2/14/2023.D4: batch # unk.This is an analysis for a photo submitted to ethicon for evaluation.During the visual analysis, the following was observed: the video shows a surgery when the stapler is removed from the trocar.Unfortunately, the video does not provide sufficient evidence to determine the root cause and the damage reported in the seal.Practical analysis should provide the necessary evidence to confirm the root cause.Based on the photo review, the event describe is not confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.We value the opportunity to fully analyze the instrument upon its return.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.An evaluation of the manufacturing documentation could not be completed as the lot number was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASX BLADELESS TROCAR 12MM
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key16194397
Report Number3005075853-2023-00376
Device Sequence Number17227826
Product Code GCJ
UDI-Device Identifier10705036003007
UDI-Public10705036003007
Combination Product (Y/N)N
Initial Reporter CountryNL
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTB12LT
Device Catalogue NumberTB12LT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/02/2023
Supplement Date Received by Manufacturer01/25/2023
Initial Report FDA Received Date01/18/2023
Supplement Report FDA Received Date02/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-