| Model Number |
860343 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The customer reported that they¿ve gone to switch it on the device with a patient in situ, and the belt has thrown this person backwards and treadmill gone into reverse.The patient is harmed.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information available and results of additional analysis by the rse, with confirmation from the manufacturer, the treadmill does not go in reverse.The customer has been instructed not to start treadmill while the patient is standing on it, per the ifu.The investigation identified no design issue or malfunction.The device remained and use with precaution instructions.
|
| |
|