• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 OSSEOTI MULTIHOLE 58MM G; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 OSSEOTI MULTIHOLE 58MM G; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bacterial Infection (1735)
Date of Event 12/20/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient underwent a first stage revision of the left total hip arthroplasty approximately 1 month post implantation due to a bacterial infection.All components were removed.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: 650-1064 ¿ ceramic taper ¿ 3064989, 650-1055 ¿ ceramic head ¿ 2020011173, 47249009700 ¿ tear drop guide wire ¿ 65316177, 166069 ¿ arcos screw ¿ 6828663, 166069 ¿ arcos screw ¿ 7032015, 166070 ¿ arcos screw ¿ 6970413, 166068 ¿ arcos screw ¿ 7053131, 166067 ¿ arcos screw ¿ 7197916.Report source: report source australia.Product will not be returned to zimmer biomet for the investigation as the product was discarded.The investigation is in process.Once the investigation as been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2023 - 00092, 0001825034 - 2023 - 00093, 0001825034 - 2023 - 00094, 0001825034 - 2023 - 00095, 0001825034 - 2023 - 00096, 0001825034 - 2023 - 00097, 0001825034 - 2023 - 00098.
 
Event or Problem Description
No additional event information to report at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: g3; h2; h3; h4; h6.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A radiograph was provided and reviewed by a health care professional.As the complaint is infection related the x-ray did not enhance the investigation.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.No device problem was found.Review of the sterile certifications were reviewed and found to be conforming with no applicable deviations.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issues identified affecting implant safety or effectiveness, implanted products are not identified as the source or contributing to the reported infection.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 OSSEOTI MULTIHOLE 58MM G
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16195424
Report Number0001825034-2023-00091
Device Sequence Number20326407
Product Code LPH
UDI-Device Identifier00887868356056
UDI-Public(01)00887868356056(17)320301(10)7196541
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K140669
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number110010267
Device Lot Number7196541
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/20/2022
Supplement Date Received by Manufacturer04/10/2023
Initial Report FDA Received Date01/18/2023
Supplement Report FDA Received Date04/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
Patient Weight185 KG
-
-