| Model Number |
799-EMVP-04 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Communication or Transmission Problem (2896); Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device displayed "off set self check failure - 1174", "internal comm failure - 1175", "internal comm failure -1474", and "compressor failure - 1002" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to a zoll approved service provider for evaluation.The customer's report was observed and attributed to the compressor.The compressor was replaced to resolve the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|