| Model Number |
PDV60FA |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Quest has not yet received the suspect device from st.Joseph's hospital.A review of the device history record revealed no quality or manufacturing concerns for the lot noted.Quest will file a supplemental report at the conclusion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported to quest medical of an alleged issue with a precision delivery iv set.The report stated that the set had a lipid line leak at the junction, just prior to the filter.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Quest received the suspect sample for investigation.It was noted that the sets were being used for lipid infusion.The lipid infusion was stopped intermittently, but the filter line was not flushed.The root cause was determined to be leaking due to lipid occlusion of the filter in the line.Quest will continue to monitor this complaint trend.Quest has concluded its investigation.
|
| |
|