• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE RF 2 PROBES, 90-S; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE RF 2 PROBES, 90-S; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0279351100
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Codes Burn(s) (1757); Superficial (First Degree) Burn (2685)
Date of Event 12/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was a patient burn.
 
Event or Problem Description
It was reported that there was a patient burn.
 
Additional Manufacturer Narrative
The device manufacturer date is not known.Alleged failure: patient burned on non-operative arm, facility thinks may have come from radiofrequency wand.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.Per visual and functional testing no defect and/or manufacturing-related condition was observed that could have contributed to the reported event.The probable root cause/s could be hot saline leaking (due to improper connection between the suction probe and the suction source), the probe suction tubing unintentionally touching the patient's skin and was maintained for a substantial amount of time.Other possibilities were activation of the probe during insertion or removal from the joint space, inadequate suction causing hot saline to leak from the joint space, and fluid irrigation.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RF 2 PROBES, 90-S
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16199930
Report Number0002936485-2023-00043
Device Sequence Number14859644
Product Code GEI
UDI-Device Identifier07613327056952
UDI-Public07613327056952
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K160050
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0279351100
Device Catalogue Number0279351100
Device Lot NumberNOT REPORTED
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/19/2022
Supplement Date Received by Manufacturer12/19/2022
Initial Report FDA Received Date01/18/2023
Supplement Report FDA Received Date03/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-