| Model Number |
7133 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: used (b)(6) 2022 as event date since the correct date was not provided.Device evaluated by mfr.Returned product consisted of an emerge mr balloon catheter.The device was microscopically and visually examined.There were numerous kinks to the hypotube.The hypotube was separated at 58cm distally from the strain relief.The separated ends of the hypotube were ovaled in shape indicating the device was kinked prior to separation.There was contrast in the inflation lumen and blood in the guidewire lumen.The balloon was tightly folded.Product analysis confirmed the reported kink as there were numerous hypotube kinks to the device.There was also an unreported hypotube separation.The separated ends were oval shaped indicating that the device was kinked prior to separation.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 10jan2023.It was reported that shaft kink occurred.A 3.00mm x 15mm fg emerge balloon catheter was selected for use.However, prior to procedure, the stent was noted to be kinked.The procedure was completed with a different device with no patient complications reported.However, returned device analysis revealed shaft detached/separated.
|
| |
|