|
The customer reported to olympus that while using the thunderbeat open fine jaw, the tissue pad burnt and detached from the tip while performing coagulation.The issue occurred during the therapeutic procedure, and the procedure was completed with a similar device.There was no patient harm associated with the event.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was not returned; therefore, the reported phenomenon or condition of the device could not be confirmed.However, it is possible to infer the cause from the results of past similar investigations.The device history record was unable to be reviewed for this device since the lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1) during the output activation in seal & cut mode, nothing was grasped in between the grasping section and the distal end of the probe (this includes after tissue resection).Therefore, the tissue pad was worn out.2) due to friction between the grasping section and the distal end of the probe, abnormal heat was generated.This might have caused the tissue pad to partially peel off.This following information is included in the instructions for use (ifu): ¿do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ olympus will continue to monitor the performance of this device.
|