• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 600023001
Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
During an svt procedure, when the amplifier was turned on, it was unable to turn from yellow to green.The case was abandoned.
 
Additional Manufacturer Narrative
Additional information: d9, g3, g6, h2, h3, h6.One ensite velocity¿ amplifier was received for evaluation at tech center.Based on the information provided to abbott and the investigation performed, the reported event was able to be confirmed, and the root cause was attributed to catheter amplifier board in slot 10.Evaluation of the logs revealed multiple occurrences of the same slot 10 post condition to channel 1.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16204102
Report Number2184149-2023-00012
Device Sequence Number5454547
Product Code DQK
UDI-Device Identifier05414734210713
UDI-Public05414734210713
Combination Product (Y/N)N
Initial Reporter CountryIN
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number600023001
Device Catalogue Number100014514
Device Lot Number6785317
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2023
Supplement Date Received by Manufacturer03/13/2023
Initial Report FDA Received Date01/19/2023
Supplement Report FDA Received Date03/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/30/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-