ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE EF 12 MM THICKNESS; PROSTHESIS, KNEE
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The returned item exhibits signs of repeated use and has one of two components disassembled / missing.The components were not returned.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.These devices are confirmed to have been a part of a previous investigation.Actions were initiated to recommend against using ultrasonic cleaning as a sterilizing technique for these devices.These devices were subsequently re-designed to account for this failure mode with a new locking mechanism.It was determined that these devices were manufactured prior to this design change.The root cause is considered to be a previously addressed design issue.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that an instrument was returned missing bearings.It is unknown when and where the device fractured, however, there was no patient involvement.Attempts have been made and all available information has been provided.
|
|