BIOTRONIK SE & CO. KG ENTOVIS SR-T PROMRI; PACEMAKER Back to Search Results
Model Number 371994
Medical Device Problem Code No Pacing (3268)
Health Effect - Clinical Code Bradycardia (1751)
Date of Event 12/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2022 patient collapsed while working out, ems claimed patient was bradycardic and patient was taken to er where the physician determined that he was not receiving pacing.The pacing lead is a competitors lead.It is unclear whether this is a lead or device issue.Pacemaker was explanted.Should additional information become available, this file will be updated.
 
Additional Manufacturer Narrative
The pacemaker was received for analysis.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this pacemaker were re-investigated and all production steps were performed accordingly.Particularly the final acceptance test proved the device functions to be as specified.Upon receipt, the pacemaker was visually inspected revealing no anomalies.In particular the header of the device was analyzed.The set screw could be easily screwed in and out, there was no foreign material inside the header bore.All dimensions of the header bore were within the range requested by the is-1 standard specifications.Also the spring element of the pacemaker did not show any deviations.At a next step the pacemaker was interrogated and the pacemakers memory content was analyzed indicating no anomalies.The battery was found to be in eri status since (b)(6) 2022 after the explantation of the device.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.There was no indication of a device malfunction.An additional long-term pacing test was initiated.During the test, each pacing pulse was recorded.The evaluation of these pacing pulses documented regular device behavior.No intermittent or permanent loss of output was present.In summary, the device is fully functional.With regard to the issues as mentioned in the complaint description, the analysis did not show any deviations from the technical specifications.The analysis did not reveal any sign of a material or manufacturing problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTOVIS SR-T PROMRI
Common Device Name
PACEMAKER
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key16205017
Report Number1028232-2023-00244
Device Sequence Number14806533
Product Code NVZ
UDI-Device Identifier04035479115197
UDI-Public(01)04035479115197(17)151231
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P950037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2015
Device Model Number371994
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2023
Supplement Date Received by Manufacturer03/10/2023
Initial Report FDA Received Date01/19/2023
Supplement Report FDA Received Date03/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/03/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age82 YR
Patient SexMale
-
-