• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE 100 ACET CUP 54MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE 100 ACET CUP 54MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-01-054
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Migration (4003)
Health Effect - Clinical Codes Cyst(s) (1800); Foreign Body Reaction (1868); Low Blood Pressure/ Hypotension (1914); Unspecified Infection (1930); Necrosis (1971); Sepsis (2067); Unspecified Tissue Injury (4559); Swelling/ Edema (4577)
Date of Event 12/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical notification received for revision due to infection.Date of implant: (b)(6) 2006, date of revision: (b)(6) 2022, (right hip).Treatment: revision; head, liner and shell were revised.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
On (b)(6)2006, patient received an uncemented right metal-on-metal total hip replacement with s-rom femur and pinnacle cup mom to address osteoarthritis.There were no reported complications.On (b)(6)2022, patient had both the right and the left total hips revised to address hematogenous sepsis, fluid collections in both hips, acute pain and synovitis of the left hip, and chronic adverse reaction to metal debris of the right hip.A total of 400 ml fluid was drained from right hip and appeared grossly infected, as well as significant evidence of adverse reaction to metal debris, including complete destruction of the abductors and posterior capsule tissues, with only necrotic tissue remaining.There were multiple pockets of pus, even down into the thigh.The greater trochanter showed some avascular damage, and the posterior wall of the acetabulum was bony deficient, allowing the well-fixed cup to have migrated neutral to slightly retroverted in position.There was mild trunnionosis of the head and stem upon head removal.The s-rom stem was well-fixed; an attempt was made to remove/revise just the stem while leaving the s-rom proximal sleeve, but the interface could not be disrupted, so both components were left in place.The acetabular cup was removed due to concerns for dislocation, and a very large cup was placed to address the cup position and stability, with polyethylene liner and new metal head.Hip was well irrigated and debrided for treatment of the infection.The patient was significantly hypotensive at this point because of the sepsis, so following closure of the right hip, arterial and central lines were placed and hemostasis restored by anesthesiologist.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not received for examination.Based on the visual analysis of the provided photographic evidence for device, it cannot be determinate a depuy's device failure or malfunction which could contribute to reported event.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE 100 ACET CUP 54MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16205024
Report Number1818910-2023-01503
Device Sequence Number5415979
Product Code LPH
UDI-Device Identifier10603295008569
UDI-Public10603295008569
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-01-054
Device Catalogue Number121701054
Device Lot NumberA6BFM1000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2023
Supplement Date Received by Manufacturer01/27/2023
03/07/2023
Initial Report FDA Received Date01/19/2023
Supplement Report FDA Received Date02/07/2023
03/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/22/2006
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP; PINNACLE 100 ACET CUP 54MM; PINNACLE MTL INS NEUT36IDX54OD; S-ROM M HEAD 36MM +3; S-ROM*SLEEVE PRX ZTT, 18D-LRG; SROM STM STD 36+12L 13X18
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexFemale
Patient Weight94 KG
Patient EthnicityHispanic
Patient RaceWhite
-
-