| Model Number |
1488020122 |
| Medical Device Problem Code |
Poor Visibility (4072)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was blurry image.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: blurry.The failure(s) identified in the investigation is consistent with the complaint record.Root cause: worn out threads most likely caused by the customer not screwing the coupler to the camera head correctly and/or wear and tear since the coupler has been on the field for almost three years without been serviced.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurry image.
|
| |
|