• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG PLUG F/TIBIAL PLATEAU; KNEE ENDOPROSTHETICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG PLUG F/TIBIAL PLATEAU; KNEE ENDOPROSTHETICS Back to Search Results
Model Number NN260P
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 07/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported to aesculap inc.That a plug f/tibial plateau (item nn260p) was implanted on (b)(6) 2013.According to the complainant, there was a gross loosening of the femoral and tibial components of the right knee.The patient underwent revision surgery on (b)(6) 2021.The adverse event/malfunction is filed under aic reference (b)(4).Associate medwatch reports: 400582818_ 2916714-2023-00001_mdr, 400582819_ 2916714-2023-00002_mdr, 400582820_ 2916714-2023-00003_mdr, 400582821_ 2916714-2023-00004_mdr.
 
Event or Problem Description
Investigation complete.
 
Additional Manufacturer Narrative
Manufacturing site evaluation: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.The device history records (dhr) were reviewed for the available lot number; the device was found to be within specification at the time of production.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUG F/TIBIAL PLATEAU
Common Device Name
KNEE ENDOPROSTHETICS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key16207523
Report Number2916714-2023-00005
Device Sequence Number14816380
Product Code JWH
UDI-Device Identifier04046963711320
UDI-Public4046963711320
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/18/2023
Report Date (Section F) 01/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2017
Device Model NumberNN260P
Device Catalogue NumberNN260P
Device Lot Number51910799
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date12/22/2022
Device Age128 MO
Event Location Hospital
Date Report to Manufacturer12/22/2022
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/19/2023
Supplement Report FDA Received Date09/18/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
Patient SexMale
Date Report Sent to FDA01/19/2023
-
-