MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number SCL-290
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported left side pain and rupture.The device has been explanted.
 
Additional Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: rupture.
 
Additional Manufacturer Narrative
Visual analysis: visual analysis of the photographs identified, rupture: observed, but cannot perform an assessment of the opening, as no microscopic analysis can be performed.Pain: unable to observe, since it is a medical event.And is not related to the device.It is not possible to determine, the lot number or catalog through the photos provided.It cannot be determined, which device is for the left or right side.Per, the photos provided.Refer to ps-qp-onev-115717: covid-19 quality plan - complaint handling.
 
Event or Problem Description
Healthcare professional reported, left side pain and rupture.The device has been explanted.
 
Event or Problem Description
Healthcare professional reported left side pain and rupture.The device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ rupture: observed, broken on the posterior side assessed as unidentified (tear) opening.(shell thickness was within specification).¿ pain: unable to observe since it is a medical event and is not related to the device.No additional observation.No further actions are required as the device was implanted.Additional, changed, and/or corrected data: b.6; d.9; h.3; h.6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key16210138
Report Number9617229-2023-00403
Device Sequence Number11875635
Product Code FTR
UDI-Device Identifier10888628029507
UDI-Public10888628029507
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/14/2021
Device Model NumberSCL-290
Device Catalogue NumberSCL-290
Device Lot Number2961087
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/06/2023
Supplement Date Received by Manufacturer02/04/2023
04/18/2023
Initial Report FDA Received Date01/19/2023
Supplement Report FDA Received Date02/24/2023
05/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/14/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexFemale
Patient Weight71 KG
-
-