| Model Number |
X060-1520 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer was made aware of a product complaint on 12/20/2022.It was reported that the distal tip of a system inserter/compressor fractured during a surgical procedure.There were no known patient complications associated with this complaint, although there was a delay in treatment to address the instrument malfunction and utilize an alternate available instrument.A return authorization was issued for return of the complaint instrument, which was received at the manufacturer for assessment on 12/27/2022.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual assessment of the returned complaint inserter/compressor showed an instrument with minimal use, as identified by crisp laser markings and a lack of surface scratches.One distal end of the inserter/compressor was fractured as reported.A functionality assessment was not performed due to the damaged condition of the inserter/compressor which was removed from distributable inventory.A dhr review was performed for the device complaint lot and there were no manufacturing anomalies identified.The device lot met all required specifications prior to being released to distributable inventory.This lot has been available for distribution since 2/10/2021.The fractured distal tip of the inserter/compressor is attached to the body of the instrument with a pin that allows for minor movement to allow for implant placement in a various patient anatomy.It may be possible that application of excessive compression force may have contributed to the instrument malfunction.There have been two other complaints of similar nature in the past 12 months.The manufacturer will continue to monitor this instrument for complaints from the field.
|
| |
|
Search Alerts/Recalls
|