| Model Number |
68LP-175 |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Patient reported deflation.Device remains implanted.This relates to the right side.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: e3.
|
| |
|
Event or Problem Description
|
|
Patient reported deflation.Device remains implanted.This relates to the right side.
|
| |
|
Search Alerts/Recalls
|