| Model Number |
168-330 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a "left side deflation".Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d6b, h6.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: deflation: observed opening sharp assessed as an unidentified (tear) opening (shell thickness was within specification).Additional observations: crease fold observed in the surface.Brown particles observed in the device.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a "left side deflation." device has been explanted and replaced.
|
| |
|