|
The device was received by nuvasive.Examination found the cage fractured into four pieces.The distal end appears to have fractured off into two pieces; additionally, the inserter attachment feature was fractured in half.The damage observed is consistent with excessive off-axis force, while attempting to distract the disc space with the interbody, insufficient space prep/anatomical obstruction in conjunction with excessive force, and/or oversized implant selection/insufficient distraction with excessive force, which would indicate procedural error as the likely cause of the event.No patient injury reported.Manufacturing review: review of the device history records notes no material non-conformances, no manufacturing errors, and no discrepancies with respect to material type, treatments, or dimensions that may have caused or contributed to this mode of failure.All devices met acceptance criteria upon release with no other complaints identified for lot: n235292.Labeling review: "warnings, cautions and precautions: the subject device is intended for use only as indicated.The implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.Correct selection of the implant is extremely important.The potential for success is increased by the selection of the proper size of the implant.While proper selection can minimize risks, the size and shape of human bones present limitations on the size and strength of implants." "notching, striking, and/or scratching of implants with any instrument should be avoided to reduce the risk of breakage." "compatibility: all implants should be used only with the appropriately designated instrument (reference surgical technique)." "pre-operative warnings: 3.Care should be used in the handling and storage of the coroent implants.The implants should not be scratched or damaged.Implants and instruments should be protected during storage and from corrosive environments.For sterile implants: assure highly aseptic surgical conditions, and use aseptic technique when removing the coroent implant from its packaging.Inspect the implant and packaging for signs of damage, including scratched or damaged devices or damage to the sterile barrier.Do not use the coroent implants if there is any evidence of damage.4.Refer to cleaning and sterilization instructions below for all non-sterile parts.5.Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient.".
|