Related manufacturer reference number: 2017865-2022-44376, related manufacturer reference number: 2017865-2022-44379.It was reported the patient presented to the emergency room due to diaphragmatic stimulation.X-ray confirmed the atrial and right ventricular leads had dislodged due to twiddlers syndrome.The atrial and right ventricular leads were successfully repositioned on (b)(6) 2022.The patient was seen for follow-up and the atrial and right ventricular leads exhibited a failure to sense and failure to capture.X-ray showed the implantable cardioverter defibrillator (icd) had migrated minimally and the atrial and right ventricular leads dislodged due to twiddlers syndrome a second time.The atrial and right ventricular leads were explanted and replaced.When attempting to connect the new leads, the set screw would not tighten and the pacing impedance was high.The icd was explanted and replaced.The patient was in stable condition.
|
The reported event of set screw issue and high pacing impedance was confirmed while the event of migration was not confirmed.The implantable cardioverter defibrillator (icd) was returned for analysis.Analysis revealed the right ventricular set screw was dislodged, stripped, and contained septum material inside the hex cavity.This material inside the hex cavity prevented full insertion of the torque driver.Additionally, septum material was found at the bottom of the right ventricular connector block and inside the lead bore.The septum material inside the lead bore prevented full insertion of the lead into the bore which resulted in high impedance measurements.The set screw anomaly was consistent with having occurred during the procedure.
|