| Model Number |
TRM-445 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv, rupture, exchange from textured to smooth due to the patient's concern of the product.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture, capsular contracture baker grade iv, and exchange from textured to smooth due to the patient's concern of the product.The device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "hole, non-specific" observed during surgery via rga.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Photo evaluation: visual analysis of the photographs identified: rupture: observed an opening the device but cannot perform an assessment of the opening as no microscopic analysis can be performed.Capsular contracture: unable to confirm as it is a medical event not related to the device.Anxiety-product/procedure: unable to confirm as it is a medical event not related to the device.It is not possible to determine the lot number or catalog through the photos provided.
|
| |
|