| Model Number |
10742326 |
| Medical Device Problem Codes |
Image Display Error/Artifact (1304); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/03/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported to siemens that a malfunction and adverse event occurred while operating the somatom force ct system.Siemens was notified that an 11 year old female patient needed to be rescanned due to image ring artifacts.As we consider children as a vulnerable patient group, this report is filed in the abundance of caution.Siemens has requested additional information in order to conduct an investigation of the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens has completed an investigation of the event.Assessment does not indicate a system failure or malfunction and no non-conformity was identified.A siemens service engineer went onsite and conducted a full evaluation of the system.Detector modules and channels were checked and found to be in good conditions.A visual inspection showed some contrast media drops on the detector.The detector was cleaned; and the sealing ring was treated with lubricant according to the documented maintenance instructions for regular service.No parts have been exchanged and the system was returned to regular operation.No further actions are to be taken as there is no negative awareness regarding the quality and performance of the system and no systematic or design issue was identified.
|
| |
|
Search Alerts/Recalls
|