| Model Number |
119112 |
| Medical Device Problem Codes |
Partial Blockage (1065); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the lumen was obstructed, and no outflow of urine could be observed after foley catheter placement.The placement of the catheter was confirmed by echography, and the tip of the catheter was placed in the bladder.The balloon was also confirmed to be inflated in the bladder.When water was injected through the lumen, no water entered the catheter, and the lumen was confirmed to have been obstructed.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the lumen was obstructed, and no outflow of urine could be observed after foley catheter placement.The placement of the catheter was confirmed by echography, and the tip of the catheter was placed in the bladder.The balloon was also confirmed to be inflated in the bladder.When water was injected through the lumen, no water entered the catheter, and the lumen was confirmed to have been obstructed.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed cause unknown.One sample exhibited the reported failure.The device had not met specifications.The reported failure is considered out of specification as the reported failure was reproduced.The product was used for patient treatment or diagnosis.The product caused the reported failure.Although a specific cause cannot be determined, based on the risk document s potential root cause for this event could be collapsed lumen and blocked lumen.A review of the dhr did not show any problems or conditions that would have contributed to the reported event.The instructions for use were found adequate and state the following: [warnings] 1.Method for use (1) do not inflate the balloon in the urethra.[the urethra may be injured.] (2) do not pull the catheter hard.[the bladder/urethra may be injured.] 2.Applicable patients ·patients with delirium who might pull out catheter [when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.] contraindications] 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) this device contains 10% povidone-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients with known allergy to silver coated catheter h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|
Event or Problem Description
|
|
It was reported that the lumen was obstructed, and no outflow of urine could be observed after foley catheter placement.The placement of the catheter was confirmed by echography, and the tip of the catheter was placed in the bladder.The balloon was also confirmed to be inflated in the bladder.When water was injected through the lumen, no water entered the catheter, and the lumen was confirmed to have been obstructed.
|
| |
|