MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE REPOSITIONABLE CLIP; SINGLE USE REPOSITIONAL CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE REPOSITIONABLE CLIP; SINGLE USE REPOSITIONAL CLIP Back to Search Results
Model Number HX-202UR
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Customer reported three events of the device being difficult to release during procedure upon regular use of the device.Procedure was therapeutic polypectomy or mucosectomy.The procedure was completed with the same device.Normal time to release the device is five minutes.Ten additional minutes were required due to this event.There is no harm or adverse impact to the patient.This type of device was used in other procedures subsequently.Opinion of the physician is that there could have been risk of perforation of the patient making the situation dangerous.
 
Manufacturer Narrative
Customer reported three events of failed devices.These events are captured in medwatches with patient identifiers (b)(6).This medwatch is for the patient identifier (b)(6).Customer has reported this event to the french national agency for the safety of medicines and health products (ansm).Customer who reported to ansm that, customer is not aware if this type of incident is provided for in the instructions for use.Customer will change manufacturer.Customer has kept the device quarantined but note returned the device as yet; as such, a definitive root cause of the reported complaint cannot be determined at this time.This event is under investigation.A supplemental report will be submitted upon completion of the investigation or upon receiving additional information.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The evaluation was completed.During device evaluation, the phenomenon was not reproduced.As there are no abnormalities as well as in appearance, we could not determine the cause of the phenomenon based on the investigation result.The clip was attached to the device when it arrived for investigation, no buckling was observed in the insertion portion, and there were no problems with slider operability.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.Based on the results of the investigation, and since no device malfunction was confirmed during evaluation, the definitive root cause of the reported issue could not be determined.It is possible that the slider was not fully pulled until the limiter placed in the control section broke.Therefore, the clip could not be released from the product.However, the exact cause could not be determined.The following information is stated in the instructions for use (ifu): ¿·operation of this instrument is based on the assumption that open surgery is possible as an emergency measure if the clip cannot be detached from the instrument or if any other unexpected circumstance takes place.In this case, refer to chapter 14, ¿emergency treatment¿.Keep the insertion portion of the instrument that extends from the biopsy valve and the insertion portion of the endoscope as straight as possible.Do not perform clipping with the insertion section being coiled.Otherwise, the force exerted on the control section may not convey to the clip adequately during clipping.This could result in reduced performance, making it impossible to close the clip or detach it from the coil sheath after clipping.·do not withdraw the instrument forcibly or change the angulation of the endoscope when the clip is grasping the tissue and is not released.Doing so may tear tissue inside the body cavity, resulting in patient injury, such as perforation, bleeding, or mucous membrane damage.Do not angulate the bending section of the endoscope or withdraw the instrument from the endoscope while the clip is grasping the tissue before being released.This could cause patient injury, such as perforation, bleeding, or mucous membrane damage.Do not try to forcibly withdraw the instrument from the endoscope if the clip cannot be detached from the instrument.Forcibly withdrawing the instrument could cause patient injury such as perforation, bleeding, or mucous membrane damage.Should the slightest irregularity and/or breakage of the parts be suspected, withdraw the instrument from the endoscope immediately according to the steps of ¿withdrawing the instrument from the endoscope¿ on page 10.Otherwise, perforation, bleeding, or mucous membrane damage may result.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE REPOSITIONABLE CLIP
Type of Device
SINGLE USE REPOSITIONAL CLIP
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16224375
MDR Text Key307908494
Report Number9614641-2023-00101
Device Sequence Number1
Product Code PKL
UDI-Device Identifier04953170385940
UDI-Public04953170385940
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K123601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2024
Device Model NumberHX-202UR
Device Lot Number12K
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-