• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9750TFX
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Bradycardia (1751)
Event Date 09/15/2022
Event Type  Injury  
Manufacturer Narrative
Per the instructions for use (ifu), conduction system defects (heart block), arrhythmias, and conduction system defects, (bradycardia, left bundle branch block (lbbb), right bundle branch block (rbbb)) which may or may not require a permanent pacemaker are potential adverse events associated with balloon aortic valvuloplasty, the use of local and/or general anesthesia, bioprosthetic heart valves, and the transcatheter heart valve (thv) procedure.Sick sinus syndrome (sss), also called sinus node dysfunction (snd), is an umbrella term for a group of abnormal heart rhythms (arrhythmias) presumably caused by a malfunction of the sinus node, the heart's primary pacemaker.Bradycardia-tachycardia syndrome is a variant of sick sinus syndrome in which slow arrhythmias and fast arrhythmias alternate.It is often associated with ischemic heart disease and valvular lesions.Sick sinus syndrome is more common in elderly adults, where the cause is often a non-specific, scar-like degeneration of the cardiac conduction system.Coronary artery disease, high blood pressure, and aortic and mitral valve diseases may be associated with sick sinus syndrome, although this association may only be incidental.Acquired snd may occur after damage to the sn artery during cardiac surgery or may be due to occlusion, such as after myocardial infarction.Another surgical cause of snd includes sn tissue damage during cannulation of the superior vena cava (svc) for cardiopulmonary bypass or extracorporeal membrane oxygenation (ecmo).Ischemic cardiac arrest may also cause snd.The natural history of snd may be highly variable, although it tends to be progressive.The only effective treatment for patients with chronic symptomatic snd is pacemaker therapy.Slow heart rates or bradycardia have many potential causes including disturbances in automaticity and conduction, medications including general anesthesia, electrolyte imbalances, advanced age, and cardiac diseases.If this cannot be managed with medication, permanent pacemaker implantation may be necessary.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest patient factors (first degree heart block) and/or procedural factors may have caused or contributed to this event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
 
Event Description
As reported by our affiliates in italy, approximately two (2) days post transfemoral transcatheter aortic valve replacement (tavr) procedure with a sapien 3 ultra valve, the patient had a permanent pacemaker implanted due to episodes of bradycardia.The event resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key16224978
MDR Text Key307913563
Report Number2015691-2023-10314
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 01/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number9750TFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/29/2022
Initial Date FDA Received01/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age84 YR
Patient SexMale
Patient Weight71 KG
-
-