Model Number 180602
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 12/29/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revised a right mako lateral uni knee due to tibial loosening.No allegations made against the implants.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Additional Manufacturer Narrative
Reported event: an event regarding loosening involving a mako baseplate was reported.The event was not confirmed.Method & results:  device evaluation and results: not performed as product was not returned.Clinician review: no medical records were received for review with a clinical consultant  .Device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there has been 1 other similar events for the lot referenced.Conclusion: the event could not be determined because insufficient information was provided.Further information such as return of the device, pre- and post-operative x-rays, primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Revised a right mako lateral uni knee due to tibial loosening.No allegations made against the implants.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCK TIBIAL BASEPLATE-LM/RL-SZ 2
Common Device Name
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
3365 enterprise ave
weston FL 33331
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16226794
Report Number3005985723-2023-00011
Device Sequence Number20190642
Product Code NPJ
UDI-Device Identifier00848486000684
UDI-Public00848486000684
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K090763
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Model Number180602
Device Catalogue Number180602
Device Lot Number26191115-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/29/2022
Supplement Date Received by Manufacturer02/28/2023
Initial Report FDA Received Date01/23/2023
Supplement Report FDA Received Date03/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/06/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age75 YR
Patient SexFemale
-
-