Section d4: serial number was requested but was not provided.Manufacturer¿s investigation conclusion: the reported event of a no external power alarm while connected to the mobile power unit (mpu) was confirmed via analysis of the submitted log file.The submitted log file contained approximately 10 days of data (b)(6) 2022 ¿ (b)(6) 2022 per the timestamp).The log file captured low voltage hazard and no external power alarms on (b)(6) 2022 from 02:45:55 to 02:55:01 due to a loss of external power while connected to the mpu.The alarms resolved once external power to the mpu was restored.The system controller backup battery supported the system during the loss of external power without any issues.No other notable alarms were observed in the log file.The alarms did not affect the controller¿s ability to operate the pump at the set speed.The mpu was not returned for analysis.Multiple attempts were made to obtain additional information, including if the mpu has a v-lock connector on the ac power cord, if the mpu ac power cord became loose from the back of the unit or from the wall outlet, if there were any environmental circumstances that would have affected ac power at the time of the event, and if the mpu was exchanged; however, no response was received.The root cause of the reported event could not be conclusively determined through this analysis.Heartmate 3 instructions for use section 7 ¿ ¿alarms and troubleshooting¿ and heartmate 3 patient handbook section 5 ¿ ¿alarms and troubleshooting¿ cover all alarms (visual and audible), including no external power and low voltage hazard alarms, and the actions to take if the issue does not resolve.Heartmate 3 instructions for use section 3 ¿ ¿powering the system¿ and heartmate 3 patient handbook section 3 ¿ ¿powering the system¿ explains how to properly use the mobile power unit (mpu) and that in the event the mpu loses power the patient is instructed to switch power sources as the backup battery in the system controller will only temporarily power the pump.Heartmate 3 instructions for use section 6 "caring for the equipment" describes how to care for and clean all equipment, including the mpu patient cable.Section 10, entitled ¿safety checklists¿, provides checklists to assist the patient in performing routine maintenance of heartmate 3 lvad, including inspecting the mpu patient cable for damage.This section also informs the user to replace any equipment or system component that appears damaged or worn.The patient handbook and the instructions for use caution the user to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|