MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 3341202
Medical Device Problem Codes Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 01/09/2016
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a 37-year-old caucasian female who underwent primary breast reconstruction with 390cc mentor memoryshape breast implants experienced bilateral capsular contracture and right sided rupture post procedure.The left sided capsular contracture was baker grade iii.The right sided capsular contracture was baker grade iv.The right sided rupture was discovered during explant.Bilateral prosthesis replacement with 430cc mentor memorygel xtra breast implants was performed on (b)(6) 2023.See 1645337-2023-00717 for contralateral prosthesis report.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture/rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The impacted product was received by the failure analysis lab on january 25, 2023.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Additional information was received on february 2, 2023.The rupture, originally reported to be right sided, was actually left sided.This report relates to the right prosthesis.Reason for device explant and/or reoperation: capsular contracture.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
The investigation of the impacted product was completed by the failure analysis lab on february 24, 2023.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During the visual evaluation, no apparent damage or visual anomalies were observed on the returned device.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYSHAPE BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key16230157
Report Number1645337-2023-00718
Device Sequence Number5273339
Product Code FTR
UDI-Device Identifier00081317005209
UDI-Public00081317005209
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P060028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/06/2020
Device Model Number3341202
Device Catalogue Number3341202
Device Lot Number6931120
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/03/2023
Supplement Date Received by Manufacturer01/25/2023
02/24/2023
Initial Report FDA Received Date01/23/2023
Supplement Report FDA Received Date02/15/2023
03/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/02/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age37 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-