| Catalog Number |
H938751 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a disconnection between the tubing and spike of a micro-volume inlet (w/large bore spike) which resulted in a leak.The issue was identified during compounding.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H4: device manufactured on october 14, 2022- october 18, 2022.H10: the device was received for evaluation.Unaided eye visual inspection was performed and the inlet spike was observed detached from the inlet tubing.The reported condition was verified.By the nature of the sample, no additional tests were performed.The cause of the condition could not be determined, however the most likely cause was due to inadequate or lack of solvent being applied to the outer diameter of the tube when it was inserted to the spike in the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|